Effect of recent severe Abutments Angulation on Implant Loaded at lower first premolar on the Fracture Resistance

Document Type : Original Article

Authors

1 Department of Fixed Prosthodontics, Faculty of Dentistry, Assuit University

2 Department of Fixed Prosthodontics, Faculty of Dentistry, EL-Ahram Canadian University, Cairo, Nahda University in Benisuief, Egypt

3 Department of Crown & Bridge department, Faculty of Dentistry, Minia University

Abstract

Background: The angulation of the abutment can affect the success of the dental implant because it controls the forces applied on the screw. Objectives: The goal of this research had been to assess the impact of abutment angulation on an implant loaded at the posterior mandible on the resultant stresses & and fracture resistance. Methods: To study the fracture resistance, the following method will be used. Fifteen casts representing missed lower first molar will be constructed & and divided into 3 groups (5 samples each) according to the type of abutments used: 1. The model consists of a Dental implant (4x 10 mm), internal hexagon straight abutment  and crown height of ten ml., 2. Dental implant (4x 10 mm), internal hexagon,15-degree angled abutment and a crown height of ten mm., 3. Dental implant (4x 10 mm), internal hexagon,30-degree angled abutment and a crown height of ten mm. The three groups will be subjected to fracture resistance tests. The data will be collected, tabulated and statistically analyzed. Results: 15 angled abutments showed the best fracture resistance (Fracture resistance test) in comparison to straight and 30 angled abutments. Conclusion: Angulation 15 abutment increases fracture resistance of the overlaying CAM-milled zirconia single crowns than straight abutment.

Highlights

Conclusions

  • Angulation 150 abutment increases fracture resistance of the overlaying CAM-milled zirconia single crowns than straight abutment.
  • Increasing the abutment angulation may lead to a deeper fracture of the restoration.

Keywords

Main Subjects


Introduction

Modern dental implants are made of biocompatible titanium and are surgically inserted into the jawbone to replace missing teeth by supporting a prosthetic tooth crown. Although implants have great long-term retention rates (approximately 95% after 5 years), failures are far more common in regions with poor bone quality and density, leading to defective patient outcomes. The majority of failures are caused by subpar clinical skill and a lack of awareness of the potentially harmful stress factors that can occur during implant installation and function (1). A dental implant's main purpose is to serve as an abutment for a prosthetic appliance that mimics the root and crown of a natural tooth (2,3). Therefore, any success criteria must include the primary support of a functional prosthesis. Additionally, the mechanical performance, Osseo-integration of the bone implant complex with low loading stresses on the bone-implant interface, & patient satisfaction with the aesthetic appearance of the implant restoration have been among the most crucial clinical criteria for prosthetic success (4,5,6). The purpose of the current investigation was to ascertain if single-unit dental implants with angled abutments may reduce stress on the surrounding bone.

 

Materials and Methods

Methods         

 I- Sample grouping

- Fifteen casts representing missed lower first molar constructed & divided into 3 groups (5 samples each) according to the type of abutments used:

1- Dental implant (4x 10 mm), internal hexagon, straight abutment and crown height of 10 mm.

2- Dental implant (4x 10 mm), internal hexagon,15-degree angled abutment & crown height of ten mm.

3- Dental implant (4x 10 mm), internal hexagon, 30-degree angled abutment & crown height of ten mm.

 

II- Model Construction:

A- Cast model preparation

2 material (acrostone), polymer and monomer were mixed till dough stage then inserted into the socket.

B- Implant insertion

The implant was loaded in the center of the acrylic resin in the socket with a ratchet until it was submerged 1 mm beneath the acrylic material's surface, before the acrostone's complete setting.

C- placement of implant Abutment

After the complete setting of the acrostone, the straight and angled abutments were placed into the 15 cast by screwing it into the implant body.

D- putty index:

A suitable metal tray size 2 was chosen and then loaded with additional silicone rubber impression material (heavy body).

After the impression material had fully set, the tray was withdrawn, loaded with a mild wash impression, and re-placed onto the mold until it had done so. The tray was then removed, the abutment was unscrewed, and the implant analogue was then inserted. Alginate was used to create an upper jaw impression using a tray size 2 and hard stone type 3.

 

III- Restoration construction

  • Core construction

Using Cercon's CAD/CAM-based techno-logies, Cercon Brain, Cercon Eye, a laser scannervand a milling machine, zirconium core thicknesses of 0.8 mm were created.

  • Crowns Construction

Using the Cercon technique, 15 crowns were machined altogether. Using a premade putty key, diagnostic wax-ups of the crowns had been completed. The putty key was filled with molten wax, which

vented out the other end. By using the putty key for each restoration, variables were decreased, and a precise replica of the wax pattern was created with full anatomical contour (7.8).

 

IV- Samples testing

The samples were subsequently put under compressive loading in a universal testing device (Shimadzu Autograph AG-50kNE, Shimadzu Co., Ltd., Japan) with a crosshead speed of one mm/minute. A steel rod with a four mm diameter had been positioned along the lower molar crown's midline fissure to apply compression force. A categorization created for the inquiry (9–11) was used to record the force (N) needed to cause the crown to fracture as well as the manner of failure.

 

V- Statistical Analysis Tests

Each group's means and standard deviations were computed and compared. The data on fracture resistance were analyzed using one-way analyses of variance (ANOVA) with accompanying post hoc Tukey pair group comparison tests(12–14) (p>0.05).

 

Results

- Fracture resistance of the tested groups the fracture resistance of the three tested groups is presented in table (1) and fig. (1).

Group B (15-degree angulated abutment) showed the highest fracture resistance 1242.41 N followed by group A (straight abutment) 1046.87. Finally, group C (30-degree angulated abutment) recorded the least fracture resistance 996.00N.

According to the Tukey test, the implant angulations considerably (p 0.05) decreased the overlaying CAM-milled zirconia single crowns' ability to withstand breakage. The samples built on the 15 angled abutments had the best fracture resistance, however, it was not statistically different from the outcomes observed for the straight abutments. In contrast, the fracture resistance of crowns bonded to the 20 angled abutments was the lowest of all the groups evaluated.

Discussion

This study examined the in vitro fracture resistance of various abutment implant angles.

 

Many abutment angulations, including 0, 15, 20, 25, 30, 35, and 50 angulations, were tried in vitro in the past (3). The majority of these investigations suggested that angula-tions greater than 30 have a negative impact on the overlaying prosthesis and the osseointegrated implant body (4). Abutment angulations have changed from 20 to zero, which is beneficial for extending the life of the overlaying prosthesis and lowering stress on the implant body and prosthesis. (7-9)

 

This study was an experiment to see which abutment angulation is better for the durability of overlaying zirconia crowns in the lower first molar area, despite the results of J. Carvello et al., (15).

 

The advantage of using biomechanical models, like the one examined in this work, is that they can give information on the clinical situation.(11.13)

 

Because of its excellent strength criteria, the Ankylos plus system (ANKYLOS-adent Gmbh Mannheim/Germany) was used in this study. According to the current findings, a 30-implant abutment angulation considerably (p 0.05) decreased the fracture resistance of the abutting CerconCAM-milled zirconia single crowns. This can be attributable to variations in the stress produced beneath the cusp points. The findings revealed that after applying pressure to the crowns, 14 of 15 cores remained undamaged. Almost half of all crown failures occurred through the midline, with the core remaining intact in all but one case while the other half of the crown was dislodged or lost.

 

 The current model system must be carefully taken into account for the general conclusion. This model was initially used for static testing of a dental material. Second, despite best efforts, it proved challenging to standardize the occlusal surface of the underlying Cercon crown due to variations in implant abutment angulation. Further investigation is required to determine the impact of the cusp angle on the scatter of compressive failure loads recorded by the model.

 

 

 

 

  •  

 

References

  • Sethi, T. Kaus, P. Sochor, D. Axmann-Krcmar, and M. Chanavaz, “Evolution of the concept of angulated abutments in implant dentistry: 14-year clinical data,” Implant Dentistry, 11, 41–51, 2002.
  • Dong, T. Kebin, C. Jiang, J. Hua, H. Wenxiu, and L. Yuyu, " A Further Finite Element Stress Analysis of Angled Abutments for an Implant Placed in the Anterior Maxilla" Computational and Mathematical Methods in Medicine, 9 Article ID 560645, 2015.
  • Clelland, K. June, C. Olivier, and A. William, "A Three-Dimensional Finite Element Stress Analysis of Angled Abutments for an Implant Placed in the Anterior Maxilla “Journal of Prosthodontics 4, 95–100, 1995.
  • Cavallaro 2, G. Gary, "studied angled implant abutments vs straight implant abutment stresses at Their implant-bone interface.؛ American Dental Association 142,150-158, 2011.
  • Cristiane 6, A. Roberto, L. Cláudio, C. Dalva, "Photoelastic analysis of stress distribution on parallel and angled implants after installation of fixed prostheses, Braz. Oral Research.18,1 2004.
  • Sakhavat 3, A. Celal, S. Faruk, A. Mehmet, T. Uğur and C. Erhan, " Effect of Abutment and Implant Shapes on Stresses in Dental Applications Using Fem" Mathematical and Compu-tational Applications, 16, 546-555, 2011.
  • Xavier E. Saab, DDS, a Jason A. Griggs, PhD,b John M. Powers, PhD,c, and Robert L. Engelmeier, DMD, MSd, (J Prosthet Dent97,85-92,2007)
  • Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent,85,585-598, 2001.
  • Brunski JB. In vivo bone response to biomechanical loading at the bone/dental implant interface. Adv Dent Res.13,99-119,1999.
  • Lin CL, WA NG JC, Chang WJ. Biomechanical interactions in tooth-implant-supported fixed partial dentures with variations in the number of splinted teeth and connector type: a finite element analysis. Clin Oral Implants Res19,107-117,2008.
  • Misch CE. Stress factors: influence on treatment planning. In: Misch CE, editor. Dental implant prosthetics. St. Louis: Elsevier Mosby;71-90,2005.
  • Juodzbalys G, Kubilius R, Eidukynas V, Raustia AM. Stress distribution in bone: single-unit implant prostheses veneered with porcelain or a new composite material. Implant Dent14, 75-166, 2005.
  • Sevimay M, Turhan F, Kilicarslan MA, Eskitascioglu G. Three-dimensional finite element analysis of the effect of different bone quality on stress distribution in an implant-supported crown. J Prosthet Dent 93,227-340, 2005.
  • John Cavallaro, Gary Greenstein, Angled implant abutments: A practical application of available knowledge, The Journal of the American Dental Association142(2):150-8 2022,